Improvement of energy resolutions in TlBr detectors
- Source :
- Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 607:112-115
- Publication Year :
- 2009
- Publisher :
- Elsevier BV, 2009.
-
Abstract
- Improvement of energy resolutions in thallium bromide (TlBr) detectors was achieved by using a gated integrator in the post amplifier stage. An energy resolution of 4.8% FWHM at 662 keV was obtained from a TlBr detector 1 mm thick at room temperature. The depth of interaction (DOI) in TlBr detectors for incident gamma-rays was determined by taking the Gaussian shaping amplifier signal to gated integrator signal ratio. The energy resolution of the detector was improved from 4.8% to 3.8% FWHM by implementation of the depth correction. The device exhibited an energy resolution of 3.3% FWHM at 662 keV by selecting gamma-ray interaction events in the detector. Long-term stability of a TlBr detector 0.36 mm thick with Tl electrode was also evaluated at room temperature. The detector operated stably for 600 h without significant performance degradations.
- Subjects :
- Physics
Nuclear and High Energy Physics
Physics::Instrumentation and Detectors
business.industry
Astrophysics::High Energy Astrophysical Phenomena
Amplifier
Detector
Signal
Semiconductor detector
Full width at half maximum
Integrator
Electrode
Optoelectronics
High Energy Physics::Experiment
business
Instrumentation
Energy (signal processing)
Subjects
Details
- ISSN :
- 01689002
- Volume :
- 607
- Database :
- OpenAIRE
- Journal :
- Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
- Accession number :
- edsair.doi...........4f6cf08cdeac4d966d852b658fb22038