Back to Search Start Over

Laser-induced plasma atomic emission spectrometry in liquid aerosols

Authors :
J. B. Simeonsson
J.D. Winefordner
Kin C. Ng
N.L. Ayala
Source :
Analytica Chimica Acta. 269:123-128
Publication Year :
1992
Publisher :
Elsevier BV, 1992.

Abstract

Atomic emission spectrometry was performed in a laser-induced plasma in air. The plasma was produced by focusing the beam of an Ar-F (193 nm) excimer laser into a liquid aerosol nebulizer system. A liquid aerosol was generated with a concentric glass nebulizer-spray chamber assembly and carried into the plasma by the argon (0.5 1 min −1 nebulizer gas. In general, the emission signal lasted ca. 35−40 μs after the laser pulse. The excitation temperature of the plasma decreased from 3994 K at 1 μs to 3607 K at 35 μs after the laser pulse. The solution limits of detection (3σ of the blank) were determined for Na, Li, In, Al, Ga, Ca, Mg, K and Sr to be 0.9, 0.3, 10, 3, 3, 8, 3, 2 and 20 μg ml −1 , respectively. The sensitivity obtained with this system is similar to that of a previous Nd: YAG wet droplet system reported in the literature.

Details

ISSN :
00032670
Volume :
269
Database :
OpenAIRE
Journal :
Analytica Chimica Acta
Accession number :