Back to Search
Start Over
Calibration of detector responses using the shape and size of band profiles
- Source :
- Journal of Chromatography A. 1089:101-104
- Publication Year :
- 2005
- Publisher :
- Elsevier BV, 2005.
-
Abstract
- A new method consisting of using band profiles to calibrate the response of a detector, without making frontal analysis measurements, is discussed. A model for the calibration-curve expression is proposed. It is convenient and reasonably accurate even when the detector response deviates slightly from linear behavior at low concentrations.
Details
- ISSN :
- 00219673
- Volume :
- 1089
- Database :
- OpenAIRE
- Journal :
- Journal of Chromatography A
- Accession number :
- edsair.doi...........4aa4bf75d6a4e7284000518511257b72
- Full Text :
- https://doi.org/10.1016/j.chroma.2005.06.077