Back to Search Start Over

Bevelled Dual-Channel Microelectrodes

Authors :
J. W. Prichard
James M. Anderson
A. Kleinhaus
L. Manuelidis
Source :
IEEE Transactions on Biomedical Engineering. :482-485
Publication Year :
1974
Publisher :
Institute of Electrical and Electronics Engineers (IEEE), 1974.

Abstract

Preparation of bevelled dual-channel glass microelectrodes is described. The bevelled tips overcome some of the problems which complicate intracellular recording. The size of the tip opening is increased, resulting in a lower resistance, which in turn confers greater current-carrying capacity and less noise susceptibility. Bevelling also preserves the sharp point, facilitating tissue penetration and minimizing neuronal trauma. The characteristics of bevelled and unbevelled electrodes filled with 0.5 or 3 M KCI, K Acetate and K 2 SO 4 were similar. Bevelled electrodes were shown to maintain a near constant resistance during current passage, while unbevelled ones showed changes in resistance that were dependent on current magnitude, polarity and duration. Such changes were caused principally by electro-osmosis.

Details

ISSN :
00189294
Database :
OpenAIRE
Journal :
IEEE Transactions on Biomedical Engineering
Accession number :
edsair.doi...........48caa98ee929b8bee885f83b3c05e2ca
Full Text :
https://doi.org/10.1109/tbme.1974.324337