Back to Search Start Over

Real-Time US-18FDG-PET/CT Image Fusion for Guidance of Thermal Ablation of 18FDG-PET-Positive Liver Metastases: The Added Value of Contrast Enhancement

Authors :
Stefano De Beni
S. Nahum Goldberg
Tiziana Ierace
Giovanni Mauri
Nicolò Gennaro
Luigi Solbiati
Marcello Rodari
Source :
CardioVascular and Interventional Radiology. 42:60-68
Publication Year :
2018
Publisher :
Springer Science and Business Media LLC, 2018.

Abstract

To assess the feasibility of US-18FDG-PET/CT fusion-guided microwave ablation of liver metastases either poorly visible or totally undetectable with US, CEUS and CT, but visualized by PET imaging. Twenty-three patients with 58 liver metastases underwent microwave ablation guided by image fusion system that combines US with 18FDG-PET/CT images. In 28/58 tumors, 18FDG-PET/CT with contrast medium (PET/CECT) was used. The registration technical feasibility, registration time, rates of correct targeting, technical success at 24 h, final result at 1 year and complications were analyzed and compared between the PET/CT and PET/CECT groups. Registration was successfully performed in all cases with a mean time of 7.8 + 1.7 min (mean + standard deviation), (4.6 + 1.5 min for PET/CECT group versus 10.9 + 1.8 min for PET/CT group, P