Back to Search Start Over

Studying the counting rate capability of a glass multigap resistive plate chamber at an increased operating temperature

Authors :
V. A. Gapienko
E. Usenko
O. P. Gavrishchuk
A. A. Semak
S. Ya. Sychkov
Yu M. Sviridov
M. N. Ukhanov
A. A. Golovin
Source :
Instruments and Experimental Techniques. 56:265-270
Publication Year :
2013
Publisher :
Pleiades Publishing Ltd, 2013.

Abstract

It is shown that a multigap resistive plate chamber made of commercial float glass is capable of sustaining high counting rates at an increased operating temperature. Two glass chambers were investigated on the test beamline of the U-70 accelerator at the Institute for High Energy Physics. The required radiation flux density at the detector was produced by means of radioactive sources. The time resolution of 80 ps or better was attained at a rate of ∼20 kHz/cm2 and an operating temperature of 45°C.