Back to Search Start Over

Implementation of electric dental handpieces by sliding mode sensorless control of PMSM

Authors :
Ming-Yang Cheng
Chun-Chia Tsao
Mi-Ching Tsai
Wen-Chun Chi
Sheng-He Wang
Source :
ISIE
Publication Year :
2013
Publisher :
IEEE, 2013.

Abstract

The ability to sustain a smooth-running practice is very important in dentistry. Because of this reason, Electric handpieces driven by permanent magnetic synchronous motors (PMSMs) have been increasingly applied in dentistry and have replaced air-driven ones due to their high efficiency and ability to sustain constant speed while operating. Additionally, they produce little noise compared to the high pitched whine of their air-driven counterparts. As is well known, the rotor position is the necessary information for driving a PMSM. Unfortunately, a dental handpiece requires repeated heat sterilization that will break down the Hall sensors, and further, lacks the space to install an incremental encoder. Therefore, sensorless speed control of a PMSM is the key technology for an electric dental handpiece. In this paper, a sliding mode observer is used to implement an electric dental handpiece. Moreover, a prototype of the electric handpiece system has been designed and built. Experimental results are presented to verify the feasibility.

Details

Database :
OpenAIRE
Journal :
2013 IEEE International Symposium on Industrial Electronics
Accession number :
edsair.doi...........40c8f37da4e5464431df8dbfcd7046ec
Full Text :
https://doi.org/10.1109/isie.2013.6563644