Back to Search Start Over

A method for comparing degradation of boron trifluoride and helium detectors in neutron and gamma fields

Authors :
T. Qian
P. Tonner
W.J.L. Buyers
N. Keller
Source :
IEEE Transactions on Nuclear Science. 45:636-642
Publication Year :
1998
Publisher :
Institute of Electrical and Electronics Engineers (IEEE), 1998.

Abstract

A method developed to measure the degradation of neutron detectors in neutron and gamma fields has been applied to five models of boron trifluoride (BF/sub 3/) detectors from major suppliers, and to a special helium (/sup 3/He) detector model. The detectors tested all have about the same nominal neutron sensitivity and overall dimensions. The showed widely different neutron and gamma durability for BF/sub 3/ models, an undesirable time-dependent gamma degradation followed by recovery for some BF/sub 3/, models, and very robust performance of the modified /sup 3/He detector.

Details

ISSN :
15581578 and 00189499
Volume :
45
Database :
OpenAIRE
Journal :
IEEE Transactions on Nuclear Science
Accession number :
edsair.doi...........3cc13ad9f296b4163811e9f9673bfce9
Full Text :
https://doi.org/10.1109/23.682463