Disposable microfluidic blood cuvette for measuring hemoglobin concentration

Authors :
Ki Tae Park
Kieth Bong Nahm
Young Ho Seo
Eui Yul Choi
Byeong Hee Kim
Won Hee Lee
Jin Kug Son
Hyung Jin Kim
Source :
Microsystem Technologies. 20:499-504
Publication Year :
2013
Publisher :
Springer Science and Business Media LLC, 2013.

Abstract

This paper presents a new air-bubble free microfluidic blood cuvette for the measurement of hemoglobin concentration. The microfluidic blood cuvette was filled with blood samples by capillary force, and hemoglobin levels in the blood were determined by measuring absorbance at the wavelength of 530 nm. Two different microfluidic blood cuvettes with dual and single sidewall microchannels were investigated. The microfluidic blood cuvette was fabricated using a polymethyl methacrylate substrate and a dry film photoresist. During the blood-filling process, air was trapped in the dual-sided wall-type cuvettes, while no air trapping occurred in the single sidewall-type cuvettes. The sensitivity of the hemoglobin measurements was more linear in a 105 μm deep microchannel than in a 35 μm deep microchannel.

Details

ISSN :
14321858 and 09467076
Volume :
20
Database :
OpenAIRE
Journal :
Microsystem Technologies
Accession number :
edsair.doi...........3ae4f0dc9acbbba13e92f60374a540a5
Full Text :
https://doi.org/10.1007/s00542-013-1954-1