Back to Search Start Over

MIT fusion γ‐ray diagnostic development

Authors :
T. Bernat
J. W. Coleman
D. H. Lo
K. W. Wenzel
R. D. Petrasso
C. K. Li
E. Hsieh
J. R. Lierzer
Source :
Review of Scientific Instruments. 63:4840-4842
Publication Year :
1992
Publisher :
AIP Publishing, 1992.

Abstract

In conjunction with {gamma}-ray diagnostic development for the NOVA Upgrade and Alcator C-MOD experiments, we have assembled a comprehensive set of absolutely calibrated {gamma}-ray sources and several {gamma}-ray detectors. These tools will be used in characterizing and calibrating new {gamma}-ray diagnostic systems as they are assembled. The {gamma}-ray sources comprise both radioisotopes and nuclear reactions. The latter are generated in the MIT Cockcroft--Walton fusion-product generator. The detectors include several standard'' NaI(Tl) scintillators and a high-resolution germanium detector. This paper briefly discusses the diagnostics planned for NOVA Upgrade and Alcator C-MOD. In addition it describes the {gamma}-ray sources and detector characterizations we have performed in the laboratory.

Details

ISSN :
10897623 and 00346748
Volume :
63
Database :
OpenAIRE
Journal :
Review of Scientific Instruments
Accession number :
edsair.doi...........38758ccb6ef3db09aa845e4aee88a12d