Back to Search Start Over

Magnet system of the 500 MHz NMR spectrometer at the Francis Bitter National Magnet Laboratory: I. Design and development of the magnet

Authors :
D. J. Ruben
L. J. Neuringer
E.S. Bobrov
W. G. Harrison
R. Weggel
J. E. C. Williams
Source :
Review of Scientific Instruments. 52:649-656
Publication Year :
1981
Publisher :
AIP Publishing, 1981.

Abstract

A high field nuclear magnetic resonance spectrometer has been put into operation at the MIT, Francis Bitter National Magnet Laboratory. The spectrometer is built around a high homogeneity magnet, the design field of which was 11.74 T, and which in fact operates in persistent mode at a field of 11.71 T. The magnet is wound using niobium–titanium superconductor cooled to 2.3 K and is housed in a special dewar designed to permit continuous subatmospheric cryogenic operation and to minimize helium consumption. This and a companion paper describe the analysis, design, and construction of the magnet.