Back to Search
Start Over
Large scale test system for upgraded DO calorimeter electronics
- Source :
- Proceedings of 1994 IEEE Nuclear Science Symposium - NSS'94.
- Publication Year :
- 2002
- Publisher :
- IEEE, 2002.
-
Abstract
- The planned upgrade of the Tevatron Collider at Fermilab will provide a higher luminosity and shorter bunch crossing than the present conditions, requiring a significant upgrade of the readout electronics for the DO Uranium-Liquid Argon Calorimeter. To study the performance of the new electronics a large scale test system (4608 electronic channels) has been built. Currently the system is instrumented with the old readout electronics. We report on the performance of the system; the developed software for data acquisition and data analysis and preliminary measurements of the new preamplifiers. >
Details
- Database :
- OpenAIRE
- Journal :
- Proceedings of 1994 IEEE Nuclear Science Symposium - NSS'94
- Accession number :
- edsair.doi...........33904ef771f7fe8237b99ec09fceac69
- Full Text :
- https://doi.org/10.1109/nssmic.1994.474520