Back to Search
Start Over
A high sensitivity and strong stability miniaturized optically-pumped magnetometer
- Source :
- Japanese Journal of Applied Physics. 61:070904
- Publication Year :
- 2022
- Publisher :
- IOP Publishing, 2022.
-
Abstract
- In this paper, we focus on the phase noise in optically-pumped magnetometer systems and investigate its sources, and use dual-phase lock-in technique to solve the problem. The results show that the phase noise is maximum when the system operates under zero-magnetic-field condition, and the main source comes from the environment stray magnetic field. The dual-phase lock-in technique improves the phase noise rejection by a factor of 10 compared to the single-phase lock-in technique. The sensitivity of the system is 22 fT Hz−1/2 and the bandwidth is 158 Hz, which offers a favorable outlook for use in clinical settings.
Details
- ISSN :
- 13474065 and 00214922
- Volume :
- 61
- Database :
- OpenAIRE
- Journal :
- Japanese Journal of Applied Physics
- Accession number :
- edsair.doi...........305e271a8b44575d0dd773de7a7c7d9e