Back to Search Start Over

Peripheral neutron dose estimation: comparison between experimental measurements and TPS estimation

Peripheral neutron dose estimation: comparison between experimental measurements and TPS estimation

Authors :
M. Ortiz-Seidel
M. T. Garcia-Hernandez
Roberto Bedogni
Francisco Sánchez-Doblado
L. Irazola
J.A. Terrón
Beatriz Sánchez-Nieto
Source :
IFMBE Proceedings ISBN: 9783319193861
Publication Year :
2015
Publisher :
Springer International Publishing, 2015.

Abstract

A newly TPS algorithm, implemented in Pinnacle3, has been developed for peripheral neutron dose estimation for radiotherapy patients. The script gives doses in several organs according to gender, treatment location and delivered high energy monitor units. In order to validate these estimations, doses for a total of 119 patients were calculated with Pinnacle3 TPS script, and compared to the experimental measurements with the new Thermal Neutron Digital Detector (TNRD). The studied patients, cover a wide range of pathologies for three different linacs. The comparison shows that with the implemented script we obtain a good correlation between measurement and theoretical values. On the other hand, periodic neutron characterization of the facility should be considered for better estimations.

Details

ISBN :
978-3-319-19386-1
ISBNs :
9783319193861
Database :
OpenAIRE
Journal :
IFMBE Proceedings ISBN: 9783319193861
Accession number :
edsair.doi...........2f750cd61dcf7ffd3fc74d0f911a2452
Full Text :
https://doi.org/10.1007/978-3-319-19387-8_96