Back to Search Start Over

Regression-based sinogram replacement for CT metal artifact reduction

Authors :
Harry E. Martz
Seemeen Karimi
Source :
Developments in X-Ray Tomography XII.
Publication Year :
2019
Publisher :
SPIE, 2019.

Abstract

In x-ray computed tomography (CT) scanning, the presence of metal objects gives rise to artifacts. Although dual-energy CT scanning and decomposition can reduce metal artifacts, in practice, the decomposition is unstable in the presence of noise and yields residual and new artifacts. A common practice in metal artifact reduction (MAR) algorithms is to use a prior-image as a guide to estimating the underlying data that are corrupted by the metal. We have developed a method in which one prior-image can be used to correct the various sinograms generated by dual-energy decomposition. We applied our method to data acquired on a commercial CT scanner. Compared to the uncorrected images, the MAR images have superior uniformity in known uniform regions while preserving edges, and better visual definition of structures.

Details

Database :
OpenAIRE
Journal :
Developments in X-Ray Tomography XII
Accession number :
edsair.doi...........2f090583c5994c53d41cc3102cf8b33f
Full Text :
https://doi.org/10.1117/12.2531682