Deformable registration of 4D computed tomography data

Authors :
George T.Y. Chen
Eike Rietzel
Source :
Medical Physics. 33:4423-4430
Publication Year :
2006
Publisher :
Wiley, 2006.

Abstract

Four-dimensional radiotherapy requires deformable registration to track delivered dose across varying anatomical states. Deformable registration based on B-splines was implemented to register 4D computed tomography data to a reference respiratory phase. To assess registration performance, anatomical landmarks were selected across ten respiratory phases in five patients. These point landmarks were transformed according to global registration parameters between different respiratory phases. Registration uncertainties were computed by subtraction of transformed and reference landmark positions. The selection of appropriate registration masks to separate independently moving anatomical subunits is crucial to registration performance. The average registration error for five landmarks for each of five patients was 2.1 mm. This level of accuracy is acceptable for most radiotherapy applications.

Details

ISSN :
00942405
Volume :
33
Database :
OpenAIRE
Journal :
Medical Physics
Accession number :
edsair.doi...........2effe98035005fbd1202641a3f748996
Full Text :
https://doi.org/10.1118/1.2361077