A study on the [18F]FDG production efficiency of the MINItraceā„¢ cyclotron in Shin Kong Memorial Hospital

Authors :
Rong-Jiun Sheu
Yuan-Hao Liu
Jui-Chang Chang
Kuo-Wei Yin
Shiang-Huei Jiang
Source :
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 562:1064-1067
Publication Year :
2006
Publisher :
Elsevier BV, 2006.

Abstract

The Cyclotron Center of Shin Kong Memorial Hospital is the most active [ 18 F]FDG producer in Taiwan. Any improvement of the production efficiency of the MINItraceā„¢ cyclotron is valuable. A theoretical activity of [ 18 F]FDG has been calculated from the proton stopping power and the interaction cross-section of the 18 O(p,n) 18 F reaction to estimate the efficiency of the production process. A high-sensitivity moderated 3 He neutron detector was installed to investigate the real-time condition of the target irradiation by monitoring the associated neutron production. Based on the results of this study, a graphical user interface program has been developed for operators to arrange the operation conditions and to analyze the performance of the production process.

Details

ISSN :
01689002
Volume :
562
Database :
OpenAIRE
Journal :
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Accession number :
edsair.doi...........2aa331c5a9c48ea5757075a518172f2f
Full Text :
https://doi.org/10.1016/j.nima.2006.02.096