Back to Search Start Over

Fabrication and characterization of a PDMS–glass hybrid continuous-flow PCR chip

Authors :
Ji Youn Lee
Seunghwan Lee
Jeong Ah Kim
Jae Jeong Kim
Shimyoung Seong
Seung Hwan Cha
Tai Hyun Park
Source :
Biochemical Engineering Journal. 29:91-97
Publication Year :
2006
Publisher :
Elsevier BV, 2006.

Abstract

A continuous-flow (CF) microchip was fabricated using polydimethylsiloxane (PDMS)/glass laminate for efficient polymerase chain reaction (PCR). The temperatures of the three separate zones, corresponding to the three PCR thermal cycling steps were controlled using an external PID temperature controller. A PCR mixture was introduced to the CF microchip using a syringe pump. The hydrophobic surface of the PDMS channel which adsorbs enzymes was modified by adding 2.5% polyvinylpyrrolidone (PVP) to the PCR mixture. The effect of flow rate in the CF chip on DNA amplification was examined and its PCR efficiency was compared with that of conventional PCR. The microchip requires only 8–30 min to produce detectable amounts of PCR products as determined by slab gel separation, whereas a conventional thermocycler requires 90–150 min over 30 amplification cycles to achieve the same result.

Details

ISSN :
1369703X
Volume :
29
Database :
OpenAIRE
Journal :
Biochemical Engineering Journal
Accession number :
edsair.doi...........2632de3791030c41c3921d5f18842455
Full Text :
https://doi.org/10.1016/j.bej.2005.02.032