Back to Search
Start Over
Measurement of helium temperatures by TVO-sensors under magnetic fields
- Source :
- Cryogenics. 42:127-131
- Publication Year :
- 2002
- Publisher :
- Elsevier BV, 2002.
-
Abstract
- Information about tests of arbitrary chosen 20 TVO resistive temperature sensors in the range from 1.5 to 5 K under magnetic fields up to 9 T is presented. Based on the obtained experimental data – more than 2000 points, the approximate and accurate methods to calculate the magneto-resistance and corresponding temperature shift due to the magnetic field are reported as well. As shown, TVO is a sensor with predictable characteristics within the investigated ranges.
Details
- ISSN :
- 00112275
- Volume :
- 42
- Database :
- OpenAIRE
- Journal :
- Cryogenics
- Accession number :
- edsair.doi...........25a67cf5731a94d657ba0407d8d40296
- Full Text :
- https://doi.org/10.1016/s0011-2275(02)00016-4