Back to Search Start Over

Flame Photometric Determination of Sodium in Primary Standard Potassium Iodate

Authors :
Ichiro Toyama
Keiichi Takauchi
Source :
Journal of the Spectroscopical Society of Japan. 7:22-25
Publication Year :
1958
Publisher :
The Spectroscopical Society of Japan, 1958.

Abstract

Flame photometric determination of 0.001% Na in primary standard potassium iodate by applying the “ Linear Extraporation Method ” is reported.By reason of the working curve of Na in 1% KIO3 solution being linear up to 1 p.p.m. Na content and its background flat, the above method becomes applicable.Under the condition of oxygen pressure 13 psi, acetylene pressure 3.5 psi, slit width 0.02 mm, and by using Beckman DU flame photometer with photomultiplier, 0.001% Na in KIO3 was determined with good precision. If sufficient care is taken to prevent the intrusion of Na, this method is more simple and more rapid as compared with the usual method.