Back to Search Start Over

Determination of Plasma Density and Temperature from Intensity Ratios of He I Emission Lines in Diversified Plasma Simulator (DiPS)

Authors :
Y.-W. Kim
J.-S. Yoon
S. J. Yoo
J.-J. Do
Y. H. Jung
K.-S. Chung
B. Lee
T. Lho
Hyun Jong Woo
Source :
Contributions to Plasma Physics. 46:460-465
Publication Year :
2006
Publisher :
Wiley, 2006.

Abstract

The intensity ratios of He I emission lines are used to determine electron density and temperature in Diversified Plasma Simulator (DiPS). The Hel line ratios are measured by optical emission spectroscopy and are calculated using the collisional-radiative equilibrium (CRE) model. The measured He I line ratios are 706.52 nm / 728.13 nm for electron temperature and 728.13 nm / 667.82 nm for electron density determination, respectively. The results obtained from the collisional-radiative equilibrium model are crosschecked with those of electrical probe measurement.

Details

ISSN :
15213986 and 08631042
Volume :
46
Database :
OpenAIRE
Journal :
Contributions to Plasma Physics
Accession number :
edsair.doi...........11e1bc27af089586d77aa4a74f44d3a0