Back to Search Start Over

Amperometric Immunosensor for Prostate Specific Antigen Based on Gold Nanoparticles/Ionic Liquid/Chitosan Hybrid Film

Authors :
Jiehua Lin
Xiujiang Pang
Chunyan He
Kong-Cheng Hu
Source :
Analytical Letters. 44:908-921
Publication Year :
2011
Publisher :
Informa UK Limited, 2011.

Abstract

The hybrid film of gold nanoparticles/ionic liquid-chitosan (GNP/IL-Ch) was first prepared by a simple one-step synthesis and used as an efficient immobilization matrix to fabricate an immunosensor. The GNP/IL-Ch film not only prevents the leakage of the IL units, but also produces a well-defined voltammetric signal due to the synergistic effects of the IL units and GNPs. By immobilizing an alkaline phosphatase (ALP)-labeled antibody in GNP/IL-Ch film, a sensitive amperometric immunosensor has been developed for prostate specific antigen (PSA). Under the optimized conditions, the immunosensor exhibits a linear range from 1.0 to 80 ng mL−1 of PSA.

Details

ISSN :
1532236X and 00032719
Volume :
44
Database :
OpenAIRE
Journal :
Analytical Letters
Accession number :
edsair.doi...........102b96c513f7f050e7364d342de69a80
Full Text :
https://doi.org/10.1080/00032711003790049