Back to Search Start Over

Biamperometric Determination of Tetracycline in Pharmaceuticals

Authors :
Renato S. Lima
Everaldo Paulo de Medeiros
Glimaldo Marino
Valberes B. Nascimento
Mário César Ugulino de Araújo
Francisco Fernandes Gambarra Neto
Wellington da Silva Lyra
Source :
Analytical Letters. 40:3070-3079
Publication Year :
2007
Publisher :
Informa UK Limited, 2007.

Abstract

A fast, reliable, and low cost biamperometric flow‐injection method, with an error of 1.3% and an analytical throughput of 55 samples h−1, for determination of tetracycline hydrochloride in pharmaceuticals capsules is proposed. The analytical curve was linear (r=0.998) in the range 10 to 50 mg l−1 using Fe(CN)6 3− and NaOH solutions as reagent and carrier stream/supporting electrolyte, respectively. A relative standard deviation of 1.6% (10 sequential injections of 30.0 mg l−1) was verified with detection and quantification limits of 0.6 and 3.4 mg l−1, respectively.

Details

ISSN :
1532236X and 00032719
Volume :
40
Database :
OpenAIRE
Journal :
Analytical Letters
Accession number :
edsair.doi...........0dbd12b339199cfc41e72c5904e0bc31
Full Text :
https://doi.org/10.1080/00032710701646115