Single electron counting using a dual MCP assembly

Authors :
Shulin Liu
Baojun Yan
Luping Yang
Kaile Wen
Peiliang Wang
Yang Yu
Ming Qi
Y. H. Yang
Xiang-Cui Lei
Tianchi Zhao
Yuekun Heng
Source :
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 830:438-443
Publication Year :
2016
Publisher :
Elsevier BV, 2016.

Abstract

The gain, pulse height resolution and peak-to-valley ratio of single electrons detected by using a Chevron configured Microchannel Plate (MCP) assembly are studied. The two MCPs are separated by a 280 µm gap and are biased by four electrodes. The purpose of the study is to determine the optimum bias voltage arrangements for single electron counting. By comparing the results of various bias voltage combinations, we conclude that good performance for the electron counting can be achieved by operating the MCP assembly in saturation mode. In addition, by applying a small reverse bias voltage across the gap while adjusting the bias voltages of the MCPs, optimum performance of electron counting can be obtained.