Back to Search Start Over

Design and fabrication of a COP-based microfluidic chip: Chronoamperometric detection of Troponin T

Authors :
María Pilar Marco
Francisco Javier del Campo
Jens Ducrée
Llibertat Abad
Francesc Xavier Muñoz
Vanessa Escamilla-Gómez
Luis J. Fernández
Susana Campuzano
Daniel Calavia
Robert Gorkin
Gloria Colom
Juan Pablo Salvador
Berta Esteban-Fernández de Ávila
María Pedrero
José M. Pingarrón
Neus Godino
Source :
ELECTROPHORESIS. 33:3187-3194
Publication Year :
2012
Publisher :
Wiley, 2012.

Abstract

This work demonstrates the design and fabrication of an all cyclo-olefin polymer based microfluidic device capable of capturing magnetic beads and performing electrochemical detection in a series of gold electrodes. The size of chip is of a microscope slide and features six independent measuring cells for multianalyte detection purposes. The aim of this work is to show that rapid prototyping techniques can be instrumental in the development of novel bioassays, particularly in clinical diagnosis applications. We show the successful determination of troponin-T, a cardiac disease marker, in the clinically relevant range of 0.05-1.0 ng/mL. This methodology achieves a detection limit of 0.017 ng/mL in PBS solutions, and is capable of detecting less than 1 ng/mL in a 1:50 human serum dilution.

Details

ISSN :
01730835
Volume :
33
Database :
OpenAIRE
Journal :
ELECTROPHORESIS
Accession number :
edsair.doi...........0ca720b2686bc63caf9b0a5bbcf078d9
Full Text :
https://doi.org/10.1002/elps.201200225