Assessment of Oxidative Stress in Patients with an Isolated Traumatic Brain Injury Using Disposable Electrochemical Test Strips

Authors :
Leonard T. Rael
David Bar-Or
Matthew M. Carrick
Raphael Bar-Or
Michael Kelly
Source :
Electroanalysis. 27:2567-2573
Publication Year :
2015
Publisher :
Wiley, 2015.

Abstract

The barriers to using plasma redox potential as a measure of oxidative stress in patients are the use of specialized equipment, necessity of sterilization between tests, and inconsistent electrode surfaces and cell geometry. These problems are resolved using disposable test strips with built in electrochemical cells. Measuring the electrode potential at a small current yields a stable value that is ∼50 mV higher for isolated traumatic brain injury (ITBI) patients than for healthy individuals. Even with natural variations within blood chemistry, this simple method may be clinically useful for assessing the degree of oxidative stress and injury severity in ITBI patients.