Low-Energy Electron Detection by Total-Absorption Cherenkov Spectrometer

Authors :
E. I. Malinovskii
A. V. Koltsov
V. I. Alekseev
A. I. L’vov
V. D. Dronov
Yu. F. Krechetov
V. V. Polyanskii
V. A. Baskov
Source :
Bulletin of the Lebedev Physics Institute. 46:289-293
Publication Year :
2019
Publisher :
Allerton Press, 2019.

Abstract

The results of calibration of the total-absorption Cherenkov spectrometer based on TF-1 lead glass of a thickness 14.8X0 using a quasi-monochromatic electron beam are presented. It is found that the energy resolution is 89–10% in the electron beam energy range E = 6–285 MeV, correspondingly.