Enzyme immobilization using a commercial kit: determination of metabolic parameters in animal blood employing a multicommutation flow system

Authors :
Pires,Cherrine Kelce
Reis,Boaventura Freire dos
Source :
Química Nova, Volume: 28, Issue: 3, Pages: 414-420, Published: JUN 2005, Química Nova v.28 n.3 2005, Química Nova, Sociedade Brasileira de Química (SBQ), instacron:SBQ
Publication Year :
2005
Publisher :
Sociedade Brasileira de QuĂ­mica, 2005.

Abstract

Automatic flow procedures based on the multicommutation concept, dedicated to the determination of 3-hydroxybutyrate, glucose and cholesterol are proposed. The enzymes were immobilized on glass beads and packed into mini-columns that were coupled to a flow system. Sampling throughputs of 55, 40 and 40 determinations per hour, linear response from 10 to 150, 50 to 600, 25 to 125 mg L-1, detection limits of 1.5, 14 and 4 mg L-1 and relative standard deviations of 1, 2 and 2% for 3-hydroxybutyrate, glucose and cholesterol, respectively, were achieved.