Back to Search
Start Over
System of active regulation of the temperature and temperature gradient in nuclear magnetic resonance sensors
- Source :
- SCOPUS00204412-1998-41-3-SID27544480488, SCOPUS00328162-1998-41-3-SID0032069009
- Publication Year :
- 1998
-
Abstract
- An active system of temperature regulation and temperature gradient in a sample on the example NMR sensor is described in this work. By supporting the given average temperature, the system of temperature-stabilization allows one to set and supervise the size and direction of the temperature gradient along a sample. The temperature of a sample is maintained with on accuracy of about ±0.1°C in a temperature interval of-150 to +200°C with a longitudinal temperature gradient no more than ±0.2°C. The value of the temperature gradient is established in the range of 0° to 10°C/cm in both directions. © 1998 MAHK Hayka/Interperiodica Publishing.
Details
- Database :
- OpenAIRE
- Journal :
- SCOPUS00204412-1998-41-3-SID27544480488, SCOPUS00328162-1998-41-3-SID0032069009
- Accession number :
- edsair.dedup.wf.001..062f5dbaf4b7251f3aac054b197e592a