Back to Search Start Over

Late complication of a Greenfield filter associating caudal migration and perforation of the abdominal aorta by a ruptured strut

Authors :
Dabbagh, A.
Chakfe, N.
Kretz, J.G.
Demri, B.
Nicolini, P.
Fuentes, C.
Mettauer, B.
Epailly, E.
Muster, D.
Eisenmann, B.
Source :
Journal of Vascular Surgery; August 1995, Vol. 22 Issue: 2 p182-187, 6p
Publication Year :
1995

Abstract

We report the case of a 67-year-old woman who was admitted for surgical removal of a Greenfield filter that had been inserted 7 years before because of recurrent deep vein thrombosis associated with pulmonary embolism. This complication appeared on a plain abdominal radiogram that showed a 7 cm distal migration of the filter, a 30-degree angulation, and rupture of a strut at the level of the hub. Computed tomography, aortography, and ascending cavography demonstrated that the inferior vena cava was perforated by the struts and that the ruptured strut had penetrated the infrarenal aorta. As demonstrated by scanning electron microscopy, the fracture was due to a structural defect of the strut at its insertion point within the hub, with no sign of corrosion. Energy-dispersive radiography analysis failed to demonstrate impurity in the metal composition. (J VASC SURG 1995;22:182-7.)

Details

Language :
English
ISSN :
07415214 and 10976809
Volume :
22
Issue :
2
Database :
Supplemental Index
Journal :
Journal of Vascular Surgery
Publication Type :
Periodical
Accession number :
ejs9933953
Full Text :
https://doi.org/10.1016/S0741-5214(95)70115-X