Clinical Evaluation of Bentley 10 Plus Bubble Oxygenator

Authors :
Basha, J. W.
Sternlieb, J.J.
Bjork, V.O.
Gabrielson, T. W.
Hodges, V.
Basha, J. W.
Sternlieb, J.J.
Bjork, V.O.
Gabrielson, T. W.
Hodges, V.
Source :
Journal of Extra-Corporeal Technology; March 1988, Vol. 20 Issue: 1 p44-47, 4p
Publication Year :
1988

Abstract

This is the first clinical report evaluating the accuracy, safety and operative simplicity of the new BENTLEY 10 PLUS (B10+) bubble oxygenator to control pO2and pCO2independently. Eleven patients undergoing cardiopulmonary bypass (CPB) were randomly selected. The Bentley Gas Stat (BGS) was used to obtain blood gas values during CPB at five-minute intervals, (N = 270). BGS calibration values were recorded. Confirming blood gas samples were sent to the hospital laboratory at approximately 15-minute intervals to ensure accuracy of the BGS. Adjustments to the gas flow controller (GFC) and total gas flow (TGF) were made to maintain the pO2and pCO2with predetermined ranges, 100–150 torr and 38–42 torr, respectively. The number of adjustments required to maintain values within ranges, GFC settings, TGF settings, Hemoglobin, and visual signs of hemolysis (hemoglobin-urea) were charted.