Back to Search
Start Over
Human Immunodeficiency Virus Type 1 (HIV-1) Plasma Load Discrepancies between the Roche COBAS AMPLICOR HIV-1 MONITOR Version 1.5 and the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 Assays
- Source :
- Journal of Clinical Microbiology; October 2007, Vol. 45 Issue: 10 p3436-3438, 3p
- Publication Year :
- 2007
-
Abstract
- ABSTRACTWe compared plasma viral load values obtained with COBAS AMPLICOR human immunodeficiency virus type 1 (HIV-1) MONITOR version 1.5 and with COBAS TaqMan HIV-1 assays. Mean values were 4.2 and 2.9 log10copies/ml, respectively, showing the lack of agreement between the two assays.
Details
- Language :
- English
- ISSN :
- 00951137 and 1098660X
- Volume :
- 45
- Issue :
- 10
- Database :
- Supplemental Index
- Journal :
- Journal of Clinical Microbiology
- Publication Type :
- Periodical
- Accession number :
- ejs57784814
- Full Text :
- https://doi.org/10.1128/JCM.00973-07