Back to Search
Start Over
201Tllabelled T1C1 dosimetry revisited
- Source :
- Nuclear Medicine Communications; February 1993, Vol. 14 Issue: 2 p104-107, 4p
- Publication Year :
- 1993
-
Abstract
- A review of the literature pertaining to the human dosimetry of 201Tl-labelled thallous chloride provides a spectrum of organ absorbed doses associated with this radiopharmaceutical. A more recent article details human quantitative studies up to 216 h postadministration. This multiorgan radiopharmacological study, however, reports the kidney dosimetry to be significantly lower than previous values; 0.0647 mGy MBq-1(0.238 rad mCi-1) compared to 0.326 mGy MBq-1(1.2 rad mCi-1). Because of the latter discrepancy the dosimetry was recalculated using the reported pharmacodynamic data. This recalculation resulted in a kidney radiation dose comparable to previously published reports of 0.514 mGy MBq-1(1.89 rad mCi-1). Also, revised absorbed doses to the various segments of the gastrointestinal tract (GIT) were significantly higher.
Details
- Language :
- English
- ISSN :
- 01433636 and 14735628
- Volume :
- 14
- Issue :
- 2
- Database :
- Supplemental Index
- Journal :
- Nuclear Medicine Communications
- Publication Type :
- Periodical
- Accession number :