Back to Search Start Over

A Study of Stresses in Alumina Universal Heads of Femoral Prostheses

Authors :
Fessler, H
Fricker†, D C
Source :
Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine; March 1989, Vol. 203 Issue: 1 p15-34, 20p
Publication Year :
1989

Abstract

Three-dimensional, frozen-stress photoelasticity was used to study the best shape for a proposed alumina universal head loaded on to a Vitallium cone taper spigot with a 30° inclined force, as in vivo. Typical cone taper friction values were reproduced in the photoelastic models.The location of the highest tensile stresses in the Mark I shape with a flat crown was found to be on the inner surface of the crown. Changing to a torispherical surface in the Mark II shape reduced this magnitude. However, the Mark III shape with a hemispherical inner crown surface gave even lower stress there, equal to the maximum value of the hoop stress at the taper, which was measured to be fairly uniform both around and along the taper except at the ends where contact pressure concentrations were found to occur and it became reduced.Lamé axisymmetric cylinder stress predictions were found to be useful approximations to measured values and were generally overestimates of the tensile hoop stress at the head taper surface.

Details

Language :
English
ISSN :
09544119
Volume :
203
Issue :
1
Database :
Supplemental Index
Journal :
Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
Publication Type :
Periodical
Accession number :
ejs48005053
Full Text :
https://doi.org/10.1243/PIME_PROC_1989_203_004_01