Back to Search
Start Over
Electromagnetic Interference from Left Ventricular Assist Device (LVAD) Inhibiting the Pacing Function of an Implantable Cardioverter-Defibrillator (ICD) Device
- Source :
- Case Reports In Cardiology; 2018, Vol. 2018 Issue: 1
- Publication Year :
- 2018
-
Abstract
- There is an increasing prevalence of patients with concomitant implantable cardioverter-defibrillators (ICDs) and left ventricular devices (LVADs). The potential for negative interactions between these continually evolving technologies is a valid concern. Previously reported interactions include inappropriate ICD therapy and interference with ICD telemetry function. Understanding the nature of such interactions and developing a comprehensive strategy to approach such situations are important. In this report, we describe a case of electromagnetic interference from LVAD inhibiting the pacing function of an ICD that was corrected by reprograming the device. We would encourage investigators to review patients with ICD and LVAD in their institutions in order to help assess the frequency and nature of these and other interactions.
Details
- Language :
- English
- ISSN :
- 20906404 and 20906412
- Volume :
- 2018
- Issue :
- 1
- Database :
- Supplemental Index
- Journal :
- Case Reports In Cardiology
- Publication Type :
- Periodical
- Accession number :
- ejs46650522
- Full Text :
- https://doi.org/10.1155/2018/6195045