Calorimetric determination of the absorbed dose‐to‐water beam quality correction factor kQfor high‐energy photon beams

Authors :
Vatnitsky, Stanislav M.
Siebers, Jeffrey V.
Miller, Daniel W.
Source :
Medical Physics; November 1995, Vol. 22 Issue: 11 p1749-1752, 4p
Publication Year :
1995

Abstract

A method has been developed for measuring photon beam quality correction factors kQusing direct transfer with a water calorimeter. kQvalues were measured for beam qualities varying between 6 and 23 MV for a Capintec PR‐06 and a PTW W30001 cylindrical ionization chamber. Measured values were intercompared with published sets of computed kQvalues and agreement was found to be within measurement uncertainty (1%, one standard deviation).