Back to Search Start Over

Point-of-Care Cepheid Xpert HIV-1 Viral Load Test in Rural African Communities Is Feasible and Reliable

Authors :
Moyo, Sikhulile
Mohammed, Terence
Wirth, Kathleen E.
Prague, Melanie
Bennett, Kara
Holme, Molly Pretorius
Mupfumi, Lucy
Sebogodi, Philemon
Moraka, Natasha O.
Boleo, Corretah
Maphorisa, Comfort N.
Seraise, Boitumelo
Gaseitsiwe, Simani
Musonda, Rosemary M.
van Widenfelt, Erik
Powis, Kathleen M.
Gaolathe, Tendani
Tchetgen Tchetgen, Eric J.
Makhema, Joseph M.
Essex, Max
Lockman, Shahin
Novitsky, Vladimir
Source :
Journal of Clinical Microbiology; October 2016, Vol. 54 Issue: 12 p3050-3055, 6p
Publication Year :
2016

Abstract

ABSTRACTRoutine monitoring of HIV-1 RNA or viral load (VL) in patients on antiretroviral therapy (ART) is important, but there are multiple impediments to VL testing in resource-constrained settings. An accurate point-of-care (POC) HIV-1 VL test could alleviate many of these challenges. We compared the performance of the Cepheid Xpert HIV-1 VL assay against the laboratory-based Abbott m2000sp/m2000rt assay (Abbott assay). ART-naive individuals participating in the Botswana Combination Prevention Project in 20 communities provided EDTA-blood specimens during household surveys. Both the POC Xpert HIV-1 VL and Abbott assays were performed on specimens sampled from 277 individuals. We found a high correlation between the Xpert HIV-1 VL and Abbott assay results (r2= 0.92; P< 0.001). The overall mean difference in the HIV-1 RNA values obtained by Xpert HIV-1 VL assay and Abbott assay was 0.34 log10copies/ml (95% confidence interval [CI], 0.26 to 0.40 log10copies/ml) (P< 0.001). Using a clinically relevant level of 1,000 copies/ml as a threshold, agreement was 90.6% (95% CI, 87.9 to 93.1%), with a sensitivity of 98.6% (95% CI, 97.2 to 100%). The two methods agreed on their detectability of HIV-1 RNA (>40 copies/ml) at 97.1% (95% CI, 95.5 to 98.7%), with a sensitivity of 99.6% (95% CI, 97.2 to 100%). The POC Cepheid Xpert HIV-1 VL assay showed high agreement and accuracy with a laboratory-based method of HIV-1 RNA testing. The POC Xpert HIV-1 VL assay tended to overestimate HIV-1 VL, although the difference was below a clinically relevant threshold of 0.5 log10copies/ml. The POC Cepheid Xpert HIV-1 VL assay is a promising tool for monitoring patients on ART in southern Africa.

Details

Language :
English
ISSN :
00951137 and 1098660X
Volume :
54
Issue :
12
Database :
Supplemental Index
Journal :
Journal of Clinical Microbiology
Publication Type :
Periodical
Accession number :
ejs40542479
Full Text :
https://doi.org/10.1128/JCM.01594-16