Back to Search
Start Over
High-resolution full-field optical coherence tomography with a Linnik microscope
- Source :
- Applied Optics; February 2002, Vol. 41 Issue: 4 p805-812, 8p
- Publication Year :
- 2002
-
Abstract
- We describe an original microscope for high-resolution optical coherence tomography applications. Our system is based on a Linnik interference microscope with high-numerical-aperture objectives. Lock-in detection of the interference signal is achieved in parallel on a CCD by use of a photoelastic birefringence modulator and full-field stroboscopic illumination with an infrared LED. Transverse cross-section (en-face, or XY) images can be obtained in real time with better than 1-µm axial (Z) resolution and 0.5-µm transverse (XY) resolution. A sensitivity of ∼80 dB is reached at a 1-image/s acquisition rate, which allows tomography in scattering media such as biological tissues.