Back to Search
Start Over
THE PNEUPAC HYPERBARIC VARIANT HB: A VENTILATOR SUITABLE FOR USE WITHIN A ONE-MAN HYPERBARIC CHAMBER
- Source :
- BJA: British Journal of Anaesthesia; October 1991, Vol. 67 Issue: 4 p488-488, 1p
- Publication Year :
- 1991
-
Abstract
- We describe the performance of the pneuPAC hyperbaric variant HB, a ventilator designed for use in a one-man hyperbaric chamber. The ventilator delivered minute volumes of 11–23 litre at 1 atm abs to 7.6–16 litre at 2.5 atm abs. The delivered minute volume may be controlled easily from outside the chamber by manipulation of the ventilator rate. 11.5–31 b.p.m.
Details
- Language :
- English
- ISSN :
- 00070912 and 14716771
- Volume :
- 67
- Issue :
- 4
- Database :
- Supplemental Index
- Journal :
- BJA: British Journal of Anaesthesia
- Publication Type :
- Periodical
- Accession number :
- ejs35816746
- Full Text :
- https://doi.org/10.1093/bja/67.4.488