New Calibration Method for Gas Chromatographic Assay of Carbon Monoxide in Blood
- Source :
- Journal of Analytical Toxicology; July 1993, Vol. 17 Issue: 4 p193-193, 1p
- Publication Year :
- 1993
-
Abstract
- The calibration method described is based on the in situ formation of carbon monoxide in a closed system by the reaction of hot concentrated sulfuric acid with formic acid. Carbon monoxide is released from hemoglobin by treatment with 85% phosphoric acid, diluted 1:2 (v/v). Carbon monoxide is analyzed by means of headspace gas chromatography on a Porapak Q 80–100 column, following catalytic reduction to methane, using a flame-ionization detector. The method was validated by comparing the results obtained for blood samples from patients suffering from carbon monoxide poisoning and previously analyzed by means of spectrophotometry.