Back to Search Start Over

Fabrication and In vivo Evaluation of Poly(3,4-ethylenedioxythiophene) Stimulus Electrodes for Fully Implantable Retinal Prosthesis

Authors :
Kigure, Chikashi
Naganuma, Hideki
Sasaki, Yuichiro
Kino, Hisashi
Tomita, Hiroshi
Tanaka, Tetsu
Source :
Japanese Journal of Applied Physics; April 2013, Vol. 52 Issue: 4 p04CL03-04CL05, 3p
Publication Year :
2013

Abstract

The development of poly(3,4-ethylenedioxythiophene) (PEDOT) stimulus electrodes and the relationship between the electrical stimulation of a rabbit retina and electrically evoked potential (EEP) were studied in detail. We fabricated implantable flexible cables with Pt, IrOx, and PEDOT electrodes and evaluated the electrochemical impedances (EIs) and charge injection capacities (CICs) of such electrodes. From the result, we confirmed that PEDOT electrodes have both lower EIs and larger CICs than Pt and IrOxelectrodes. In addition, we performed in vivo experiments with PEDOT electrodes and clarified the relationships between the electrical stimulation of the rabbit retina and EEP. It is highly probable that visual restoration will be realized safely with PEDOT electrodes.

Details

Language :
English
ISSN :
00214922 and 13474065
Volume :
52
Issue :
4
Database :
Supplemental Index
Journal :
Japanese Journal of Applied Physics
Publication Type :
Periodical
Accession number :
ejs30152123
Full Text :
https://doi.org/10.7567/JJAP.52.04CL03