Reconsideration of Total Erythrocyte Destruction Phenomenon

Authors :
Tayama, Eiki
Shimono, Takatsugu
Makinouchi, Kenzo
Ohtsubo, Satoshi
Nakazawa, Tadashi
Takami, Yoshiyuki
Niimi, Yoshinari
Glueck, Julie
Nosé, Yukihiko
Source :
Artificial Organs; July 1997, Vol. 21 Issue: 7 p704-709, 6p
Publication Year :
1997

Abstract

Abstract: During a particular long‐term in vitro hemolysis test, the plasma free hemoglobin suddenly increased even though the hemolysis level had risen linearly for the previous several hours. This phenomenon was dubbed the total destruction of erythrocytes (TDE) phenomenon, and it was hypothesized that this was the result of the accumulation of sublethal damage to erythrocytes. It was suggested that the TDE might demonstrate the hemolytic characteristics of a pump more sensitively than a conventional hemolysis test. However, the previous report did not consider the effects of temperature or contamination. To study these effects, 3 long‐term hemolysis tests were concluded under the following conditions. For Study 1 blood temperature was maintained at 27°C (n = 2); for Study 2, at 37°C (n = 4); and for Study 3, at 37°C with gentamicin (n = 4). The BioMedicus and Nikkiso pumps were used as they were in our previous report. Gas sterilization of all circuits and pumps preceded experimentation. In Studies 1 and 3, hemolysis increased linearly for 29 h. However, in Study 2 a sudden increase of hemolysis occurred for both pumps. Possible causes of this were the dramatic changes in environmental factors such as severe acidosis, high 0, and glucose consumption, and CO2accumulation. In contrast, neither Study 1 nor Study 3 showed a sudden increase in hemolysis. The plasma free hemoglobin increased linearly in both groups until 29 h of pumping. The environmental changes resulting from contamination were considered to be the cause of the sudden increase in hemolysis. In conclusion, the TDE did not reflect mechanical blood cell damage, but rather different environment situations. Hemolysis increased linearly up to 29 h in either 27°C or germ‐free conditions.

Details

Language :
English
ISSN :
0160564X and 15251594
Volume :
21
Issue :
7
Database :
Supplemental Index
Journal :
Artificial Organs
Publication Type :
Periodical
Accession number :
ejs24997803
Full Text :
https://doi.org/10.1111/j.1525-1594.1997.tb03727.x