Improving contrast and tracking of tags in cardiac magnetic resonance images

Authors :
Koerkamp, M.A.M. Groot
Snoep, G.
Muijtjens, A.M.M.
Kemerink, G.J.
Source :
Magnetic Resonance in Medicine; May 1999, Vol. 41 Issue: 5 p973-982, 10p
Publication Year :
1999

Abstract

Cardiac tagging permits non‐invasive study of myocardial motion with high accuracy. Unfortunately, tagging contrast is impaired at later heart phases due to longitudinal relaxation. Histogram modification is presented as a method for improving contrast in later, faded images of a tagging series by altering these images such that their intensity histograms approximate the shape of the first, unfaded image of the series. This technique greatly improves the contrast and facilitates automatic detection of tags. Furthermore, a method is described for automatically tracking tag positions in short‐axis images of the left ventricle modulated with a tagging grid. The method differs from previously reported methods in that, in one single filtering process in the Fourier domain, both the grid crossings as well as the grid centers are detected, and thus increased sampling resolution is obtained. The method was validated by applying a mathematical model of left ventricular motion to tagged images of the thigh muscle. The average discrepancy between theoretically predicted and automatically detected tag positions was 0.04 ± 0.17 mm (mean ± SD). Magn Reson Med 41:973–982, 1999. © 1999 Wiley‐Liss, Inc.

Details

Language :
English
ISSN :
07403194 and 15222594
Volume :
41
Issue :
5
Database :
Supplemental Index
Journal :
Magnetic Resonance in Medicine
Publication Type :
Periodical
Accession number :
ejs24565757
Full Text :
https://doi.org/10.1002/(SICI)1522-2594(199905)41:5<973::AID-MRM17>3.0.CO;2-8