Back to Search
Start Over
Analysis of Trace Levels of Volatile Organic Contaminants in Municipal Drinking Water by Glass Capillary Gas Chromatography Using Simultaneous Flame Ionization and Electron Capture Detection
- Source :
- Analytical Letters; January 1980, Vol. 13 Issue: 16 p1417-1429, 13p
- Publication Year :
- 1980
-
Abstract
- A glass capillary gas chromatographic system using simultaneous flame ionization (FID) and electron capture (ECD) detection has been employed for the analysis of trace volatile organic pollutants in a municipal drinking water supply. The use of dual detectors with glass capillary columns allows resolution and detection of both halogenated and non-halogenated compounds simultaneously at less than microgram per liter (ppb) concentrations. By using diethyl ether as a solvent in preparing standard solutions of volatile organic compounds, standardization is made more accurate due to a reduction in solvent masking of early eluting peaks of interest. Additionally, ether shows promise for use in an internal standard method for quantification of VOA. These techniques were found to alleviate problems previously encountered in the analysis of purgeable organics and are described.
Details
- Language :
- English
- ISSN :
- 00032719 and 1532236X
- Volume :
- 13
- Issue :
- 16
- Database :
- Supplemental Index
- Journal :
- Analytical Letters
- Publication Type :
- Periodical
- Accession number :
- ejs11086070
- Full Text :
- https://doi.org/10.1080/00032718008081378