Back to Search Start Over

Comparison of fluoroscopic operator eye exposures when working from femoral region, side, or head of patient.

Authors :
Jordan Ray, M.
Mohammad, Fawzi
Taylor, William B.
Cura, Marco
Savage, Clare
Source :
Baylor University Medical Center Proceedings; Jul2013, Vol. 26 Issue 3, p243-246, 4p, 2 Black and White Photographs, 1 Chart, 1 Graph
Publication Year :
2013

Abstract

Operator radiation exposure is an important occupational hazard compounded over the course of an interventional radiologist's career. This study compared operator radiation dose to the eye and head for different positions around the patient. Compared with cases performed from the femoral region, exposures were 1.8 times higher at the side, and 1.6 times higher at the head, using conventional aprons, table shields, and mobile suspended shields. Exposures were 99% lower when using a suspended personal radiation protection system in all positions. In conclusion, standing at the side or head results in higher head exposures in a conventional setup. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
08998280
Volume :
26
Issue :
3
Database :
Supplemental Index
Journal :
Baylor University Medical Center Proceedings
Publication Type :
Academic Journal
Accession number :
88386923
Full Text :
https://doi.org/10.1080/08998280.2013.11928971