Mitigating the risk of overdosing TMS due to coil-to-scalp distance: An electric field modeling study.

Authors :
Caulfield, Kevin A.
LaPorta, Samantha M.
Walton, Rhiannon M.
Collins, Elisabeth V.
Summers, Philipp M.
Cho, Jennifer Y.
Antonucci, Michael U.
Opitz, Alexander
George, Mark S.
McTeague, Lisa M.
Source :
Brain Stimulation; Jul2024, Vol. 17 Issue 4, p970-974, 5p
Publication Year :
2024

Abstract

• We report a patient with different coil-to-scalp distances over motor (11mm) and prefrontal (2mm) regions due to hairstyle. • 120 % MT dosing would have resulted in prefrontal E-fields equivalent to 180 % MT. • E-field dosing may mitigate over-stimulation and determined that only 71.3 % MT was needed in this case. [ABSTRACT FROM AUTHOR]