Back to Search Start Over

Failure to defibrillate or cardiovert due to premature truncation of biphasic shocks from implantable defibrillators.

Authors :
Hauser, Robert G.
Kapphahn-Bergs, Melanie
Casey, Susan A.
Witt, Dawn R.
Sengupta, Jay D.
Source :
Heart Rhythm; Feb2024, Vol. 21 Issue 2, p143-149, 7p
Publication Year :
2024

Abstract

In 2022 and 2023, Medtronic recalled implantable defibrillators because they may deliver less than full-energy shocks. The 2022 problem truncates the second phase of the waveform (SCP-T2), resulting in ∼32-J shocks, and is mitigated by downloadable software. The 2023 malfunction truncates the first phase of the waveform, resulting in 0- to 12-J shocks due to a glassed feedthrough problem (GFT-T1) that might be avoided by programming B>AX shock polarity. The purpose of this study was to assess the consequences of GFT-T1 and SCP-T2 shocks in the Food and Drug Administration's Manufacturers and User Facility Device Experience (MAUDE) database and to estimate the incidences of GFT-T1 and SCP-T2. We analyzed MAUDE reports supplemented by Medtronic data; lead failures were excluded. The incidences of SCP-T2 and GFT-T1 were estimated using USA volumes for devices with glassed feedthroughs. One hundred thirty-two devices delivered truncated shocks: 27 (20.5%) were GFT-T1; 103 (78.0%) were SCP-T2; and 2 (1.5%) truncated both phases (BOTH-T1&2). Of 54 ventricular fibrillation (VF) patients, 21 (38.9%) were not defibrillated by truncated shocks: 8 (38.1%) received GFT-T1 shocks, 12 (57.1%) received SCP-T2 shocks, and 1 received a BOTH-T1&2 shock; 2 patients suffered unrelated deaths; 1 was externally rescued; 1 outcome was unknown; the others were defibrillated by subsequent shocks or terminated spontaneously. The majority of patients (79.1%) shocked for ventricular tachycardia (VT) were converted, primarily (94.1%) by SCP-T2 shocks. Estimated incidences of GFT-T1 and SCP-T2 were 0.0078%–0.0088% and 0.1062%–0.1110%. GFT-T1 and SCP-T2 shocks can result in failure to terminate VF/VT, but they may be preventable. Although the incidences of these truncated shocks are very low, heightened surveillance is warranted. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
15475271
Volume :
21
Issue :
2
Database :
Supplemental Index
Journal :
Heart Rhythm
Publication Type :
Academic Journal
Accession number :
174708705
Full Text :
https://doi.org/10.1016/j.hrthm.2023.11.006