Back to Search Start Over

Bioprosthetic Valve Remodeling in Nonfracturable Surgical Valves: Impact on THV Expansion and Hydrodynamic Performance.

Authors :
Meier, David
Puehler, Thomas
Lutter, Georg
Shen, Carol
Lai, Althea
Gill, Hacina
Akodad, Mariama
Tzimas, Georgios
Chhatriwalla, Adnan
Allen, Keith B.
Blanke, Philipp
Payne, Geoffrey W.
Wood, David A.
Leipsic, Jonathon A.
Webb, John G.
Sellers, Stephanie L.
Sathananthan, Janarthanan
Source :
JACC: Cardiovascular Interventions; Jul2023, Vol. 16 Issue 13, p1594-1608, 15p
Publication Year :
2023

Abstract

There are limited data on the effect of bioprosthetic valve remodeling (BVR) on transcatheter heart valve (THV) expansion and function following valve-in-valve (VIV) transcatheter aortic valve replacement (TAVR) in a nonfracturable surgical heart valve (SHV). This study sought to assess the impact of BVR of nonfracturable SHVs on THVs after VIV implantation. VIV TAVR was performed using 23-mm SAPIEN3 (S3, Edwards Lifesciences) or 23/26-mm Evolut Pro (Medtronic) THVs implanted in 21/23-mm Trifecta (Abbott Structural Heart) and 21/23-mm Hancock (Medtronic) SHVs with BVR performed with a noncompliant TRUE balloon (Bard Peripheral Vascular Inc). Hydrodynamic assessment was performed, and multimodality imaging including micro–computed tomography was performed before and after BVR to assess THV and SHV expansion. BVR resulted in limited improvement of THV expansion. The largest gain in expansion was observed for the S3 in the 21-mm Trifecta with up to a 12.7% increase in expansion at the outflow of the valve. Minimal change was observed at the level of the sewing ring. The Hancock was less amenable to BVR with lower final expansion dimensions than the Trifecta. BVR also resulted in notable surgical post flaring of up to 17.6°, which was generally more marked with the S3 than with the Evolut Pro. Finally, BVR resulted in very limited improvement in hydrodynamic function. Severe pinwheeling was observed with the S3, which improved slightly but persisted despite BVR. When performing VIV TAVR inside a Trifecta and Hancock SHV, BVR had a limited impact on THV expansion and resulted in SHV post flaring with unknown consequences on coronary obstruction risk and long-term THV function. [Display omitted] [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
19368798
Volume :
16
Issue :
13
Database :
Supplemental Index
Journal :
JACC: Cardiovascular Interventions
Publication Type :
Academic Journal
Accession number :
164865997
Full Text :
https://doi.org/10.1016/j.jcin.2023.03.035