Back to Search Start Over

CT doses based on clinical indications. New national DRL's in Iceland.

Authors :
Guðjónsdóttir, Jónína
Michelsen, Sigurbjörg Sigurðard
Björnsdóttir, Guðlaug
Guðmundsdóttir, Valdís Klara
Source :
Physica Medica; May2023, Vol. 109, pN.PAG-N.PAG, 1p
Publication Year :
2023

Abstract

• New Icelandic national diagnostic reference levels. • Data collected from all computed tomography scanners in Iceland. • National diagnostic reference levels, based on clinical indications, were established. • Computed tomography examinations have been linked to clinical indications. A diagnostic reference level (DRL) is an investigation level to use in the optimization of a medical exposure using ionizing radiation. The aim of this project was to gather dose data from computed tomography (CT) studies in Iceland with the purpose of updating existing national DRL and proposing DRLs based on clinical indications. Dose data (total dose length product (DLP) and CT dose index) were retrospectively collected from all CT scanners in Iceland for 50 patients for all common CT studies. After cleaning the data, the data set contained total DLP for 8129 patients. Considering dose-relevant parameters such as the number of phases and scan length, each CT study was assigned to one of 42 study types and (at least) one clinical indication. Data were received from seven or more scanners for 13 clinical indications and from all scanners providing coronary angiography. There was considerable variation in the median total DLP from different scanners, e.g., from 13 to 139 mGycm for Sinusitis , and, the range of individual total DLPs varied between scanners. This underscores the importance of monitoring patients' exposure. The study shows that there is room for optimization. New Icelandic NDRLs were proposed for the total DLP of CT examinations based on ten different indications, including: Brain ischemia (1060 mGycm), Interstitial lung disease (310 mGycm) and Appendicitis (690 mGycm). The proposed NDRLs were the first clinical indication based NDRLs in Iceland. The Icelandic Radiation Safety Authority has decided to update the NDRLs based on the results of this study. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
11201797
Volume :
109
Database :
Supplemental Index
Journal :
Physica Medica
Publication Type :
Academic Journal
Accession number :
163515866
Full Text :
https://doi.org/10.1016/j.ejmp.2023.102576