Balloon-Expandable Valve for Treatment of Evolut Valve Failure: Implications on Neoskirt Height and Leaflet Overhang.

Authors :
Akodad, Mariama
Sellers, Stephanie
Landes, Uri
Meier, David
Tang, Gilbert H.L.
Gada, Hemal
Rogers, Toby
Caskey, Michael
Rutkin, Bruce
Puri, Rishi
Rovin, Joshua
Leipsic, Jonathon
Sondergaard, Lars
Grubb, Kendra J.
Gleason, Patrick
Garde, Kshitija
Tadros, Hatem
Teodoru, Sebastian
Wood, David A.
Webb, John G.
Source :
JACC: Cardiovascular Interventions; Feb2022, Vol. 15 Issue 4, p368-377, 10p
Publication Year :
2022

Abstract

This study sought to determine the degree of Evolut (Medtronic) leaflet pinning, diameter expansion, leaflet overhang, and performance at different implant depths of the balloon-expandable Sapien 3 (S3, Edwards Lifesciences LLC) transcatheter heart valve (THV) within the Evolut THV. Preservation of coronary access and flow is a major factor when considering the treatment of failed Evolut THVs. An in vitro study was performed with 20-, 23-, 26-, and 29-mm S3 THVs deployed within 23-, 26-, 29-, and 34-mm Evolut R THVs, respectively. The S3 outflow was positioned at various depths at node 4, 5, and 6 of the Evolut R. Neoskirt height, leaflet overhang, performance, and Evolut R valve housing diameter expansion were assessed under physiological conditions as per ISO 5840-3 standard. The neoskirt height for the Evolut R was shorter when the S3 outflow was positioned at node 4 compared with node 6 (node 4 height for 23 mm = 16.3 mm, 26 mm = 17.1 mm, 29 mm = 18.3 mm, and 34 mm = 19.9 mm vs node 6 height for 23 mm = 23.9 mm, 26 mm = 23.4 mm, 29 mm = 24.7 mm, and 34 mm = 27 mm Evolut R). All configurations exhibited acceptable hydrodynamic performance irrespective of the degree of leaflet overhang, except the 29-mm S3 implanted in 34-mm Evolut R at node 4 (regurgitant fraction >20%). The valve housing radius of the index Evolut R increased when the S3 was implanted, with the increase ranging from 0 to 2.5 mm. Placement of the S3 at a lower implant position within an index Evolut R reduces the neoskirt height with no significant compromise to S3 valve function despite a higher degree of leaflet overhang. Low S3 implantation may facilitate future coronary access after redo transcatheter aortic valve replacement. [Display omitted] [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
19368798
Volume :
15
Issue :
4
Database :
Supplemental Index
Journal :
JACC: Cardiovascular Interventions
Publication Type :
Academic Journal
Accession number :
155154066
Full Text :
https://doi.org/10.1016/j.jcin.2021.12.021